• Ads

RTD Pt-100 Four Point Alarm

Here 4052 is used as an analog multiplexer, U3A TL062 opamp is wired as a 1mA Constant Current Source. It pumps 1mA into U2 4052 pin 13 X . Depending on BCD code on inputs A, B of 4052 the current is routed to any one of the four RTD 100E, whose one end is connected to X0-X3. The current then flows to GND thru the Platinum 100E resistance. The Y0-Y3 monitors the mV developed on RTD in tandem with X0-X3 positions. Like a ganged rotary switch. The Output mV at Y is the mV of Active Channel as selected by the BCD of 4029.

RTD Pt-100 Four Point Alarm – del90001

The BCD is provided by 4029 counter which is clocked by a 555. U5A cancels out the 100E mV (1mA * 100E) of a cold RTD and Amplifies the differential mV. This output of U5A is in proportion with measured temperature. U5B compares the mV that was obtained with a preset mV of POT1, which is a user setting called setpoint. The difference is amplified by U5B which is saturated by U3B comparator which adds a little hysteresis too. R7-C2 further dampen and slow the response. This finally drives Q2 to provide a Logic Signal indicating if temperature is below or above setpoint.

RTD Pt-100 Four Point AlarmFour points in a Closed Loop Temperature Control System is Monitored and a Alarm set to go off when the temperature goes beyond the set limit.

Sample and Hold with Standby CD4053

A, B, and C are the Digital Control for x, y and z input and output pairs.The voltage at Vinx is stored in C1 when A goes high, when A is low the voltage stored in C1 is read by buffer U2A.

It could be used in DMM circuits as Analog Memory DMM range, AC-DC mode, logic control

Sample and Hold with Standby CD4053

Another application of  Mux 8 Channel 4-20mA Analog Multiplexer

The stby or standby input should be low when sample and hold is operating. If stby is taken high then C1 Cap is isolated and leakage is minimum. The supply of +/- 7.5V is chosen as OFF resistance of 4053 is high at this supply.